| General | Details |
|---|---|
| Name | HACH LT200 Dry Thermostat User Manual |
| Make | HACH |
| Language | English |
| Filetype | PDF (Download) |
| File size | 0.46 MB |

Hach LT200 Dry Manual: Questions and Answers
Hach LT200 Dry specifications
| Power supply | Single heating block: 115 V: 300 VA; 230 V: 450 VA |
| Setpoint range | LT 200: 37 to 150 °C (98.6 to 302 °F) |
| Storage temperature | –40 to +60 °C (–40 to +140 °F) |
| Humidity | 90% (non-condensing) |
| Weight | Single heating block: 2 kg (4.4 lb) |
Can I use the Hach LT200 Dry with 16 mm tubes in a 20 mm heating block?
Yes. Put reducing adapters in the openings. User Manual, p. 9
How do I set up the Hach LT200 Dry for the first time?
Place it on a stable, level, heat-resistant surface. Connect the power cord to the back plug, then to an earthing-contact socket. Push the power switch on the back. The language screen appears on startup.
- Put the instrument on a stable, level, heat-resistant surface.
- Connect the power cord to the power plug on the back.
- Connect the cord to a power socket with earthing contact.
- Push the power switch on the back to turn on the instrument. User Manual, p. 8
How do I select the language on the Hach LT200 Dry?
The language menu appears when you first turn on the instrument.
- Hold down the left key and turn the instrument on.
- Push the right arrow key to scroll through the menu.
- When the correct language is shown, push the key for that language. User Manual, p. 8
How do I know when the Hach LT200 Dry heating block reaches the set temperature?
Two beeps sound when the heating block reaches the selected temperature. User Manual, p. 12